S1100 ICU Ventilator

The ventilator is suitable for various kinds of medical institutions for cardiopulmonary resuscitation respiratory support. Acute respiratory failure caused by various reasons or incomplete oxygenation dysfunction. Intra-operation,  post-operation respiratory support, other respiratory treatment.

Features

  • 12.1” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform. 

  • Recycle breathing pipe, ensure easy operating and tidiness.

  • Multiple working modes such as volume control and pressure limit, adapt to wide range patient.

  • Vaporizer with temperature, flow compensation and self-lock function, maintains safety anytime.

  • Multiple parameters monitoring interface, let users know the patient conditions in all aspects;

  • Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.

  • Stable, low noise air compressor, create quiet work environment for doctors.

Safety

  • Three level alarming system, visual and sound alarm information.

  • Several alarm, reminder, and protection functions.

  • Advanced power management control technology.

  • Built-in backup power source, when outside power source goes off, back-up power source starts to work.

  • Self-check before running, eliminates system error.

  • Separate design of electric and gas, keeps safe running of ventilator.

Detailed Specifications

           

Ventilation modes:

IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV,SIGH,MANU

Ventilator parameter ranges

 

Tidal volume(Vt)

0 ~ 2000mL

Frequency (Freq)

Oxygen concentration

1bpm ~ 100 bpm

21%-100%

I/E

4:1~1:8

PEEP

0cmH2O ~ 40 cmH2O

Pressure Limit

20 cmH2O ~ 100 cmH2O

Monitored Parameters

 

Frequency (Freq)

0 /min ~ 100 /min

Tidal volume(Vt)

0 mL ~ 2500 mL

Oxygen concentration

15 % ~ 100 %

 

Oscillographs display:

 

P-T(pressure – time)

 

F-T(flow – time)

 

P-V loop (pressure – volume loop)


 

Alarm and protection  

The AC power failure alarm

Power failure or no connection

Internal battery backup low voltage alarm

≤11.3±0.3V

No tidal volume

No tidal volume within 6s

High Minute Volume alarm

LowMinute Volume alarm

5L/min-99L/min

1L/min-30L/min

High Airway pressure alarm

Low Airway pressure alarm

High oxygen concentration alarm

Low oxygen concentration alarm

20cmH2O-100cmH2O

0cmH2O-20cmH2O

19%-100%

18%-99%

Continuous pressure alarm

(PEEP+1.5cmH2O) over 16s

Suffocation warning

Fan error

Oxygen deficit

                5-60s

Show on screen

Show on screen

The maximum limited pressure 

<12.5 kPa

 

 

Working conditions

 

Gas source

O2,AIR

Pressure

280kPa-600kPa

Voltage

-220V±22V

Power frequency

50Hz±1Hz

Input power

900VA(with air compressor)

250VA(without air compressor)

Size

1st Wooden case packing size(main engine):L560*W 560*H 605mm , GW:35KG ; NW:17KG

2nd Wooden case packing size(air compressor):L670*W700*H 1160 mm , GW:79KG ; NW:46.2KG